How to Calibrate a Hardness Tester

mechanical

Hardness testers measure material hardness using methods such as Rockwell, Brinell, Vickers, or Knoop. Calibration involves verifying the applied force, indenter geometry, and measurement system using certified hardness test blocks. Hardness testing is critical for heat treatment verification and incoming material inspection.

Required Reference Standards

  • ASTM E18 - Rockwell hardness testing
  • ASTM E10 - Brinell hardness testing
  • ASTM E384 - Microhardness (Vickers/Knoop) testing
  • Certified hardness test blocks for each scale in use

Calibration Procedure

  1. 1

    Visual and Mechanical Inspection

    Inspect the tester frame, anvil, and indenter for damage or contamination. Verify the test surface and anvil are clean and seated properly. Check that the indenter is the correct type for the scale being verified (diamond cone for HRC, ball for HRB, etc.).

  2. 2

    Force Verification

    If the procedure requires it, verify the applied test forces using a calibrated force-measuring device (elastic proving ring or load cell). Check both preliminary (minor) and total (major) forces for Rockwell, or the full force for Brinell/Vickers.

  3. 3

    Indenter Verification

    Verify the indenter geometry using optical measurement or by substituting a certified reference indenter. For Rockwell diamond indenters, the cone angle and tip radius must meet ASTM E18 requirements.

  4. 4

    Performance Verification with Test Blocks

    Perform five hardness impressions on each certified test block for the scales in use. Space impressions at least three diameters apart and at least 2.5 diameters from any edge. Record each reading.

  5. 5

    Error Calculation

    Calculate the average of the five readings and compare to the certified block value. The error is the difference between the average reading and the certified value. Also assess the repeatability (range of five readings).

  6. 6

    Adjustment and Retesting

    If errors exceed limits, adjust the tester per manufacturer instructions (force adjustment, zero offset, etc.) and retest. Record as-found and as-left results.

  7. 7

    Documentation

    Record all readings, test block IDs and certified values, indenter ID, and environmental conditions. Issue the calibration certificate with pass/fail determination for each scale tested.

Acceptance Criteria

Per ASTM E18 (Rockwell): the average of five readings must agree with the certified block value within ±1.0 HRC (or ±2.0 HRB). The repeatability (range of five readings) must not exceed 1.0 HRC. Similar criteria apply per ASTM E10 (Brinell) and E384 (Vickers).

Typical Calibration Interval

12 months, with daily verification using test blocks

FAQ

How often should hardness testers be verified with test blocks?

ASTM E18 recommends daily verification at the beginning of each shift and whenever the indenter or anvil is changed. Full calibration (including force and indenter checks) is performed annually. The daily verification with certified test blocks is a check, not a full calibration.

What is the difference between direct and indirect verification?

Direct verification measures the tester's forces and indenter geometry directly. Indirect verification tests the tester's performance by making impressions on certified test blocks. Both are required for full calibration per ASTM E18.

Can I use old hardness test blocks for daily verification?

Test blocks have a shelf life (typically 5 years from certification) and degrade over time. Expired test blocks should not be used for calibration or verification. Surface contamination, oxidation, and handling marks can also invalidate test blocks.

Track Hardness Tester Calibrations Automatically

CalibrationOS tracks due dates, stores certificates, and generates audit-ready reports.

Get Started Free